Nikon F/41F Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Objectifs d'appareil photo Nikon F/41F. Nikon F/41F User's Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Medical-Nikkor

Medical-Nikkor 120mm f/41F Nikon INSTRUCTION MANUAL

Page 2 - NOMENCLATURE

-OPERATION-continued-----------------------1··1.-1··1··1··1 800 400 200 100 50 2~ 10 SETTING THE ASAjlSO FILM SPEED The flash output of the ringl

Page 3 - Power Supply (Optional)

SETTING THE SHUTTER SPEED DIAL The shutter speed setting necessary for proper flash synchronization depends on the type of sync cord used, the

Page 4 - FOREWORD

- OPERATION-continued-----------------------12 READY-LIGHT INDICATION After the power supply is switched on, a red LED ready-light, built into

Page 5 - MOUNTING THE LENS

FOCUSING/DETERMINING THE EXPOSURE This lens focuses in the same way as other Nikon/Nikkor lenses do. Therefore, if the reproduction ratio is less im

Page 6 - POWER SUPPLY

- OPERATION-continued------------------------14 SETTING THE REPRODUCTION RATIO To set the reproduction ratio between 1/11 X and 1 X, simply turn

Page 7 - T mark next to the socket

• Prime Lens Only • With 2X Close-Up Attachment Lens Reproduc-Focused m Depth of field tion ratio distance (ft) mm (in.) 1/11X 1.6 + 25.4 ! + 1.00

Page 8 - * In the

-OPERATION-continued-----------------------16 IMPRINTING THE REPRODUCTION RATIO The OAT A button on the side of the lens barrel determines wheth

Page 9

STOP-DOWN EXPOSURE MEASUREMENT Although the lens has a built-in ringlight for shooting, you can take pictures without it. In this case, use

Page 10 - SETTING THE ASAjlSO FILM

-OPERATION-continued-----------------------FOCUSING LAMP If the ambient light is dim, push the LAMP button to illuminate the subject for focu

Page 11 - SETTING THE SHUTTER

RECOMMENDED FOCUSING SCREENS-------------Various interchangeable focusing screens are available for F3-and F2-series cameras to su it any type of len

Page 12 - If the Medical-Nikkor

NOMENCLATURE DATA button LED read Power socket Power cord connection mark Focusing lamp R t flash Film speed/aperture index Sync cord connection mark

Page 13

CAUTIONS---------1. Due to the electrical characteristics of the Medical-Nikkor's ringlight flash and focusing lamp, this lens sho

Page 14 -

ABOUT GROUNDING------• Before connecting or disconnecting the ground cord, make sure the LA-2's AC plug is not con-nected to a wall outlet.

Page 15

SPECIFICATIONS---------------------Focallength Maximum aperture Lens construction Picture angle Distance scale Aperture scale Diaphragm Reproduction r

Page 16 - 2X close-up

DC Power Unit LD-2 (with 8 AA-type batteries) Recycling times Number of flashes Film Manganese Alkaline-Manganese Alkaline-speed manganese manganese

Page 17 - EXPOSURE

No reproduction in any form of this booklet. in whole or in part (except for brief quotation in critical articles or reviews). may be made without wri

Page 18 - FOCUSING LAMP

Cords x ~ /& /,' ~ ~ J h' 'I, 1 tt SlS 394 312 272 2 J 191 1" IS. 131 111 liS x08 08S 09 Q9S 1 'I 12 13 IS 18 ] it 108 0

Page 19 - 0 © © © © ©

CONTENTS-----------------NOMENCLATURE ... . . ... ... 2-3 FOREWORD... . . . . ... . . . . ... 4 OPERATION

Page 20 - CAUTIONS

OPERATION---------------------------------------------MOUNTING THE LENS Position the lens in the camera's bayonet mount, lining up the bla

Page 21 - ACCESSORIES

-OPERATION-continued-----------------------CONNECTING THE POWER SUPPLY The Medical-Nikkor 120mm f/4 accepts either AC or DC power units, allowing the

Page 22

• AC Power Unit LA-2 Before using the AC unit, make sure the voltage selector is set for the proper voltage. The unit has separate settings for 220 an

Page 23 - 98mmt/>

-OPERATION-continued-----------------------8 CONNECTING THE SYNC CORD Two separate sync cords are supplied with the lens. Select the cord best suite

Page 24 - NIPPON KOGAKU K.K

• 3-Pin Sync Cord SC-22 The 3-pin sync cord can be used with all Nikon and Nikkormat cameras having standard ISO-type hot shoe. When the S